i::scan photometer

i::scan

the new miniature multi-parameter spectrophotometer probe will revolutionize online water quality monitoring: From very cost sensitive applications down to highly resolved “Smart Water Grids”, in small unmanned plants, or even in single building protection. Overall the new i::scan combines the high performance of a multi wavelength spectrophotometer with even lower costs than simple photometers!

 

  • s::can plug & measure
  • turbidity: measurement according to EPA 180.1 and ISO 7027, 90° scattering (for a 35 mm path length)
  • new light emitting technology
  • no consumables as well as no moving parts
  • special, non-fouling optical window material
  • low power consumption (specifically less than 1 W typical)
  • dual-beam compensated optics
  • optional automatic cleaning compressed air (InSitu, only for version -075 with fixed cable) or autobrush
  • multiple versions for multiple applications
  • long term stable and also 100 % corrosion free
  • either plug connection or fixed cable
  • at least 5000 hours maintenance free operation
  • mounting and measurement directly in the media (in situ) or in flow cell
  • can be mounted directly in a mains pipe or pressure pipe
  • operation though s::can terminals & s::can Software

Visit i-scan.at for more information!(Opens in a new tab or window)

 

References - i::scan

Accessories - Installation

submersed

  • ruck::sack
    • submersible Autobrush for spectrometer probes and i::scan
    • exchangeble brushes for spectrometer probe with path length 35, 15, 5 mm and i::scan 35, 5 mm
    • one Basis module (motor unit) for all Versions
    • shelter protects the brush from clogging

     

    technical specification
    power supply: 12 VDC
    power consumption (typical): 150 mA (average)
    power consumption (max.): 300 mA
    cable length: 7.5 m
    housing material: POM-C
    dimensions (width x height x xdepth): 182 x 46 x 36.5 mm
    weight (min.): 750 g
    installation / mounting: submersed
    operating pressure 0 ... 0.5 bar
    ingress protection class: IP68
    storage temperature: -20 ... 80 °C
    storage humidity: 0 ... 95 %

    part numbers
    F-146-rs-35: ruck::sack submersible AutoBrush for spectro::lyser 35 mm
    F-146-rs-15: ruck::sack submersible AutoBrush for spectro::lyser 15 mm
    F-146-rs-05: ruck::sack submersible AutoBrush for spectro::lyser 5 mm
    F-146-rs-iscan-35: ruck::sack submersible AutoBrush for i::scan 35 mm
    F-146-rs-iscan-05: ruck::sack submersible AutoBrush for i::scan 5 mm

    recommended accessoires
    F-146-brush-35: Brush for ruck::sack - 35 mm spectro::lyser (spare part)
    F-146-brush-15: Brush for ruck::sack - 15 mm spectro::lyser (spare part)
    F-146-brush-05: Brush for ruck::sack - 5 mm spectro::lyser (spare part)
    F-146-brush-iscan-35: Brush for ruck::sack - 35 mm i::scan (spare part)
    F-146-brush-iscan-05: Brush for ruck::sack - 5 mm i::scan (spare part)

  • i::scan carrier

    for easy attachment of one i::scan

    technical specification 
    housing material: PVC, POM-C
    dimensions (diameter x lenght): 58 x 115 (without connection fitting)
    installation / mounting: submersed

     

    part numbers
    F-110-iscan: carrier for i::scan, for easy horizontal attachment
    F-120-iscan: carrier for i::scan, for easy vertical attachment
    F-130-iscan: carrier for i::scan, for easy 45° attachment

  • fixing adapter - stainless steel
    • supports the proper and easy installation of s::can systems
    • quick mounting and demounting of InSitu installed s::can probes and sensors

     

    technical specification
    housing material: stainless steel
    weight (min.): 2.6 kg
    process connection: ID 50 mm
    part number: F-15

    recommended accessories
    F-14-3 mounting pipe - stainless steel

by-pass

  • i::scan flow cell for i::scan and up to 3 additional s::can probes
    • for proper and easy flow-through installation of one i::scan and up to three s::scan physical probes
    • automatic cleaning with autobrush for i::scan available (optional)

    technical specification
    housing material: POM-C
    dimensions (diameter x length): 106 x 103
    weight (min.): 1 kg (without autobrush)
    process connection: G 1/4", hose nozzle 7mm
    installation / mounting: flow cell
    operating temperature: 0 ... 50 °C
    operating pressure: 0 ... 6 bar
    part number: F-46-four-iscan

    recommended accessories
    F-501-eco-us System Panel micro::station US
    F-501-eco-eu System Panel micro::station EU
    F-45-process process connection 1/4" G
    F-45-fl ow-1 Automatic flow control unit
    F-45-strain Inlet strainer
    F-446-m-iscan i::scan autobrush for F-46-flow cells

  • flow cell for i::scan (waste water)
    • for proper and easy flow-through installation of one i::scan in waste water application

     

    technical specifications 
    housing material: PVC
    dimensions (lenght x width x height): 177 x 83 x 90 mm
    process connection: 1" inside thread
    installation / mounting: flow cell
    recommended flow: < 40 l/min
    part number: F-48-iscan

  • single flow-cell (by-pass fitting), Pom-C
    • for proper and easy flow-through installation of one i::scan
    • automatic cleaning with autobrush for i::scan available (optional)

     

    technical specification
    housing material: POM-C
    dimensions (diameter x length): 70 x 86
    process connection: G 1/4", hose nozzle 7mm
    installation / mounting: flow cell
    operating temperature: 0 ... 50 °C
    operating pressure: 0 ... 6 bar
    part number: F-46-iscan (without autobrush)
    part number: F-446-iscan (with autobrush)

    recommended accessories
    F-501-eco-us System Panel micro::station US
    F-501-eco-eu System Panel micro::station EU
    F-45-process process connection 1/4" G
    F-45-flow-1 Automatic flow control unit
    F-45-strain Inlet strainer
    F-446-m-iscan i::scan autobrush for F-46-flow cells

by-pass

  • Pressure mounting for i::scan in-pipe Installation (i::scan removal under pressure)
    • for proper and easy installation of one i::scan in a pressure pipe
    • under pressure drilling of pipes possible (for PE, PVC, DCI, steel and AC pipes)
    • the i::scan can be mounted and demounted under pressure without interuption of the water flow

     

    technical specification
    housing material:stainless steel
    dimensions (height): 550 m (max.)
    weight (min.): 5 kg
    process connection: Hawle shut off saddle with a O ring DN 50 - DN 300 (others on request)
    operating pressure: 0 ... 12 bar
    part number: F-160-iscan

More Accessories

  • extension cable for s::can sensors and s::can ISE probes
    • extension of sensor cable
    • IP 68 plug connection
    • flexible adjustment to local conditions

     

     technical specification

    cable length: 10, 20 or 30 m
    assembling: ex works
    housing material: jacket: PU
    protection class: IP 68
    interface connection to s::can sensors:

    part numbers
    C-210-sensor (10 m)
    C-220-sensor (20 m)
    C-230-sensor (30 m)