June 2023 I The latest generation of condu::lyser uses different communication settings (38400 Baud, 8 Bits, parity odd, 1 stop bit) than condu::lysers as shipped before May 2022.
New condu::lyser generation
condu::lyser
The latest generation of condu::lyser uses different communication settings (38400 Baud, 8 Bits, parity odd, 1 stop bit) than condu::lysers as shipped before May 2022. You can identify such condu::lysers by 38400 Baud printed onto the labels of the condu:::ylser (highlighted in green).
Main advantages of this modernization are:
• One central communication standard is established and used by all s::can sensors when connected to s::can terminals.
• Additional information can be transferred by a condu::lyser to s::can terminals (e.g. firmware version, serial number)
The following terminals are fully compatible:
• con::lyte D-320 (firmware v7.11 or higher)
• con::cube D-330
• con::line D-500